10/30 leader board - Lab Equip & Supplies Dir

ReactRaman 785

Available from METTLER TOLEDO

ReactRaman is a probe-based, Raman spectroscopy instrument that can monitor crystallization processes, and reveal reaction mechanisms and kinetics in single or multiphase systems. A compact base unit with a high-performance spectrometer and sampling technology is coupled with the next generation iC Raman™ 7.1 software. This powerful combination means that accurate results can be obtained and converted quickly into the evidence scientists need to make informed decisions.

Specifications - ReactRaman 785
For Use In Laboratory: EasyMax/OptiMax
Software iC Raman 7
Base Unit Description Laboratory base unit
Optical Range (Base Unit) 100cm-1 to 3200cm-1
Excitation Wavelength 785 nm
Excitation Power Max: 400mW; Standard: <250mW
Detector 2-stage TE Cooled CCD
Base Unit Dimension H: 77mm [3.0 in.]; W: 250mm [9.8 in.]; L: 250mm [9.8 in.]
Fiber Connections FC/PC (both excitation and collection)
Base Unit Temperature Range – 5 °C – 35 °C
Certifications CE; NRTL-C/US
Laser Classification Class IIIB laser. Compliant with 21 CFR 1040.10 and 1040.11
Power 100-240 VAC, 50/60Hz, 1.5A
Sampling Technology Fixed or Interchangeable Probe; Flowcell [optional, compatible with interchangeable Fiber Head]
Probe Wetted Material C-276, Sapphire, Gold
Probe Wetted Diameter 12.7mm [0.5 in.]; 9.5mm [0.375 in]
Probe Wetted Length 12.7mm dia: 260mm [10.25 in.]; 9.5mm dia: 248mm [9.75 in.]
Probe Temperature Range -40°C to 300°C
Probe Pressure Rating 413 barg [6,000 psig]
Probe Fiber Length 1.7m [5.6ft]

Rate and Review METTLER TOLEDO

Rating: 0 - 0 reviews
Let users know what you would have wanted to know about this company.
Click on a star to rate this company
 
  Privacy Policy.
When writing a review, please adhere to the Review Guidelines.